The purpose of this note is to compare how pyFAI
and ImageD11
treat the detector position. In particular, we derive how “PONI”
detector parameters refined with pyFAI
can be transformed into
ImageD11
parameters.
In both packages, the transformation from pixel space to 3D laboratory coordinates is carried out in 4 steps:
Transformation from “pixel space” to the “detector coordinate system”. The detector coordinate system is a 3D coordinate system centered on the (0,0) pixel of the detector.
Correction for linear offsets, i.e. the position of the (0,0) pixel relative to the beam axis.
Correction for the origin/diffractometer-to-detector distance. The sample and diffractometer center of rotation are assumed to be located at the origin.
A series of rotations for the detector coordinate system relative to the laboratory coordinates.
Unfortunately, the conventions chosen by pyFAI
and ImageD11
differ. For example, pyFAI
applies the origin-to-detector distance
correction before rotations, whereas ImageD11
applies it after
rotations. Furthermore, they employ different coordinate systems.
Detector¶
We consider a pixelated 2D imaging detector. In “pixel space”, the position of a given pixel is given by the horizontal and vertical pixel numbers, \(d_H\) and \(d_V\). We assume that looking along the beam axis into the detector, \(d_H\) increases towards the right (towards the center of the synchrotron) and \(d_V\) towards the top. For clarity, we assign the unit \(\mathrm{px}\) to these coordinates.
The pixel numbers \(d_H\) and \(d_V\) are transformed into 3D “detector” coordinates by a function \(D\):
This function will account for the detector’s pixel size and the orientation and direction of pixel rows and columns relative to the detector coordinate system. Furthermore it may apply a distortion correction. This, however, is beyond the scope of this note.
Limiting ourselves to linear functions, \(D\) takes the form of a
matrix with two columns and three rows. We will see below that the
different choices of laboratory coordinate systems yield different
\(D\)-matrices for pyFAI
and ImageD11
. We assume that the
pixels have a constant horizontal and vertical size,
\(\mathrm{pxsize}_H\) and \(\mathrm{pxsize}_V\). Both are given
in units of length per pixel. pyFAI
specifically defines the unit of
length as meter, we will therefore use pixel sizes in units of
\(\mathrm{m}/\mathrm{px}\) throughout this note.
The position and orientation of this detector relative to the laboratory coordinates are described below.
Geometry definition of pyFAI
¶
Coordinates¶
pyFAI
uses a coordinate system where the first axis (1) is
vertically up (\(y\)), the second axis (2) is horizontal (\(x\))
towards the ring center (starboard), and the third axis (3) along the
beam (\(z\)). Note that in this order (1, 2, 3) is a right-handed
coordinate system, which makes \(xyz\) in the usual order a
left-handed coordinate system!
Units¶
All dimensions in pyFAI
are in meter and all rotation are in
radians.
Parameters¶
pyFAI
describes the position and orientation of the detector by six
variables, collectively called the PONI, for point of normal incidence.
In addition, a detector calibration is provided in the PONI-file to
convert pixel coordinates into real-space coordinates. Here we limit our
discussion to the simplest case, i.e. a pixel size as discussed above.
- Rotations:
\(\theta_1\), \(\theta_2\) and \(\theta_3\) describe the detector’s orientation relative to the laboratory coordinate system.
- Offsets:
\(\mathrm{poni}_1\) and \(\mathrm{poni}_2\) describe the offsets of pixel (0,0) relative to the “point of normal incidence”. In the absence of rotations the point of normal incidence is defined by the intersection of the direct beam beam axis with the detector.
- Distance:
\(L\) describes the distance from the origin of the laboratory system to the point of normal incidence.
Detector¶
The transformation from pixel space to pyFAI
detector coordinates is
given by
Offsets¶
The PONI parameters are: a distance \(L\), the vertical (\(y\)) and horizontal (\(x\)) coordinates of the point of normal incidence in meters, \(\mathrm{poni}_1\) and \(\mathrm{poni}_2\). The inversion of the \(x\) and \(y\) axes is due to the arrangement of the detector data, with \(x\)-rows being the “slow” axis and \(y\)-columns the “fast” axis. Extra care has to be taken with the signs of the rotations when converting form this coordinate system to another.
pyFAI
applies both the offset correction and the origin-to-detector
distance after the transformation from pixel space to the detector
system, but before rotations,
Let \(L\) be the distance from the origin/sample/diffractometer center of rotation. In the absence of any detector rotations, \(L\) is taken along \(p_3\) (beam axis, \(z\)), \(p_1\) along the \(y\)-axis (vertical) and \(p_2\) along the \(x\)-axis (horizontal). Then the laboratory coordinates before rotation are
Rotations¶
The detector rotations are taken about the origin of the coordinate system (sample position). We define the following right-handed rotation matrices:
The rotations 1 and 2 in pyFAI
are left handed, i.e. the sign of
\(\theta_1\) and \(\theta_2\) is inverted.
The combined pyFAI
rotation matrix is then
which yields the final laboratory coordinates after rotation
Inversion: Finding where a scattered beam hits the detector¶
For a 3DXRD-type simulation, we have to determine the pixel where a scattered ray intercepts the detector. Let \(A\) be the scattering center of a ray within a sample volume (grain, sub-grain or voxel). The Bragg condition and grain orientation pre-define the direction of the scattered beam, \(\vec{k}\). The coordinates \(A_{1,2,3}\) and \(k_{1,2,3}\) are specified in the laboratory system.
The inversion eq. [eq-tpyFAI] is straight-forward:
The third line (\(\ldots = L\)) of eq. [eq-find-alpha] is then used to determine the free parameter \(\alpha\), which in turn is used in the first and second lines to find \(p_{1,2}\) and thus \(d_{1,2}\).
As the most trivial example we consider the case of no rotations, \(\theta_1 = \theta_2 = \theta_3 = 0\). Then
We see also that when all rotations are zero, \((\mathrm{poni}_1, \mathrm{poni_2})\) are the real space coordinates of the direct beam (\(A_{1,2,3}=k_{1,2}=0\)) .
Geometry definition of ImageD11
¶
For maximum convenience, ImageD11
defines almost everything
differently than pyFAI
.
Coordinates¶
ImageD11
uses the ID06 coordinate system with \(x\) along the
beam, \(y\) to port (away from the ring center), and \(z\) up.
Units¶
As the problem is somewhat scale-invariant, ImageD11
allows a free
choice of the unit of length, which we will call \(X\) here. The
same unit has to be used for all translations, and for the pixel size of
the detector. The default used in the code appears to be
\(X = 1\,\mathrm{\mu m}\), but it might as well be Planck lengths,
millimeters, inches, meters, tlalcuahuitl, furlongs, nautical miles,
light years, kparsec, or whatever else floats your boat. The only
requirement is that you can actually measure and express the detector
pixel size and COR-to-detector distance in your units of choice. Since
we want to compare to pyFAI
, we choose \(X=1\,\mathrm{m}\).
Rotations are given in radians.
Parameters¶
ImageD11
defines the detector geometry via the following parameters:
- Beam center:
\(y_{\mathrm{center}}\) and \(z_{\mathrm{center}}\) define the position of the direct beam on the detector. Contrary to
pyFAI
, the beam center is given in pixel space, in units of \(\mathrm{px}\).- Pixel size:
The horizontal and vertical pixel size are defined by \(y_{\mathrm{size}}\) and \(z_{\mathrm{size}}\) in \({X}/{\mathrm{px}}\). With the right choice of the unit of length \(X\), these corresponds directly to the pixel sizes \(\mathrm{pxsize}_H\) and \(\mathrm{pxsize}_V\) defined above.
- Detector flip matrix:
\(O = \begin{bmatrix} o_{11} & o_{12} \\ o_{21} & o_{22} \end{bmatrix}\). This matrix takes care ofcorrecting typical problems with the way pixel coordinates are arranged on the detector. If, e.g., the detector is rotated by \(90^{\circ}\), then \(O=\begin{bmatrix} 0 & 1 \\ -1 & 0\end{bmatrix}\). If left and right (or up and down) are inverted on the detector, then \(o_{22} = -1\) (\(o_{11}=-1\)).
- Rotations:
Detector tilts \(t_x\), \(t_y\), and \(t_z\), in \(\mathrm{rad}\). The center of rotation is the point where the direct beam intersects the detector.
- Distance:
\(\Delta\), in units \(X\), is the distance between the origin to the point where the direct beam intersects the detector. Note that this is again different from the definition of
pyFAI
.
It appears that these conventions where defined under the assumption that the detector is more or less centered in the direct beam, and that the detector tilts are small.
Transformation¶
The implementation in the code transform.py
is using the following
equations:
Note that the order of \(y\) and \(z\) is not the same in eqs. [eq-p] and [eq-tImageD11].
By combining the detector flip matrix \(O\) and the pixel size into a detector \(D\) matrix, this can be written as
Conversion¶
Assume that the same detector geometry is described by the two notations. How can the parameters be converted from one to the other?
Detector \(D\)-matrix¶
The pixel size is the same in both notations, \(y_{\mathrm{size}} = \mathrm{pxsize}_H\) and \(z_{\mathrm{size}} = \mathrm{pxsize}_V\).
As pyFAI
does not allow for detector flipping, \(o_{11}=1\),
\(o_{22}=-1\) (because the sign of the horizontal axis is inverted
between ImageD11
and pyFAI
) and \(o_{12}=o_{21}=0\). For the
detector setup described above, with \(d_V\) increasing to the top
and \(d_H\) increasing towards the center of the synchrotron
(i.e. opposite to the positive \(y\)-direction), eq. [eq-DImageD11]
becomes
Coordinates¶
Both notations use the same sign for the vertical and beam axes. The sign of the horizontal transverse axis, however, is inverted.
The transformation between the different coordinate systems is then achieved by:
where \(t_{\mathtt{ImageD11}}\) is given by eq. [eq-tImageD11], and \(t_{\mathtt{pyFAI}}\) is given by eq. [eq-tpyFAI]. The matrix \(G\) performs the change of axes (\(x \leftrightarrow z\), \(y \leftrightarrow -y\)) and has the convenient property \(G^2 = 1\).
Substituting these equations into eq. [eq-coordconv], one can them
attempt to convert pyFAI
parameters into ImageD11
parameters and
vice versa.
Rotations¶
Take an arbitrary vector \(d\) with \(d_{\mathtt{ImageD11}}
= \begin{bmatrix} a \\ b \\ c \end{bmatrix}\). We first transform this
into the pyFAI
coordinate system by multiplication with \(G\),
and then apply an arbitrary rotation matrix, once in before (in
pyFAI
coordinates, \(R_{\mathtt{pyFAI}}\)) and once after the
transformation (in ImageD11
coordinates,
\(R_{\mathtt{ImageD11}}\)).
Comparing the last two lines, we find that with
the transformation is applicable for each and any vector \(d\). Because \(G^{-1} = G\) this transformation can also be applied to a series of rotations: \(G \cdot R \cdot R' = (G \cdot R \cdot G) \cdot (G \cdot R' \cdot G) \cdot G\).
Applying this to the rotations matrices defined in eqs. [eq-rot1]–[eq-rot3] shows, unsurprisingly, that this coordinate transformation is an exchange of rotation axes \(x\) and \(y\), and a change of sign for \(y\).
Applying this transformation to the pyFAI
rotation matrix can
comparing to the ImageD11
rotation matrix, we see
We find that, by divine intervention 1 and despite all the efforts to choose incompatible conventions, the effective order of rotations is actually the same between ``ImageD11`` and ``pyFAI``. Consequently, there is a direct correspondence with only a change of sign between \(\theta_z\) and \(\theta_1\):
Translations and offsets¶
Inserting eqs. [eq-thetax]–[eq-thetaz] into [eq-transformation], we find
With a little help from our friend Mathematica, we find for the
conversion from pyFAI
to ImageD11
and for the conversion from ImageD11
to pyFAI
- 1
May his noodly appendages forever touch you!